Probe and patient
Measurement feedback
Freeze the image, choose Calipers, then click twice to place each caliper. Place one caliper 3 mm straight back from the posterior globe along the nerve, then a second caliper across the sheath, outer wall to outer wall, at that depth.
ONSD log
No measurements saved yet.
Patient and physiology
Machine
cm
Eye and orbit anatomy (transverse section)
Tap a structure name to highlight it.
The optic nerve sheath is a continuation of the dura. The subarachnoid space around the nerve communicates with the intracranial CSF, so raised ICP distends the sheath, most of all just behind the globe.
Ultrasound appearance and where to measure
- A: find the posterior globe where the optic nerve enters (the optic disc).
- A to B: go 3 mm straight back along the nerve axis, where the sheath is widest.
- At B: measure perpendicular to the nerve (and its vessels), from the outer edge of one hyperechoic sheath wall to the outer edge of the other. Never include the black shadows outside the sheath.
- Align first: rock until the nerve is parallel to the beam, fan until the sheath is widest.
- Confirm it is the nerve: turn on color Doppler briefly. Red choroidal flow arcs along the back of the globe, and the central retinal artery (red) and vein (blue) appear as a linear red and blue pair in the center of the nerve just behind the disc, the red flashing with each heartbeat. A dark stripe with no flow, especially one that widens with depth, is shadow artifact.
- Repeat in the transverse and sagittal planes, scan both eyes, and average (best three).
- If the ONSD is 5 to 6 mm, measure the optic disc elevation (ODE).